FOR ION CHROMATOGRAPHY
HYDROXAN™ COLUMNS
THE FIRST RUSSIAN-DEVELOPED PRODUCT OF ITS KIND
A Russian-made hydroxide-selective anion-exchange HPLC column designed for the qualitative and quantitative determination of trace concentrations of inorganic anions and low-molecular-weight organic acids in aqueous matrices.
COLUMN FEATURES
Production is located in St. Petersburg.

The matrix synthesis, particle functionalization, and packing of the sorbent into the column body are all developed in Russia. The microspheres are made of a highly cross-linked styrene-based polymer, providing mechanical resistance to pressures of up to 400 atm, compatibility with aqueous-organic media, and stability over a wide pH range.

The columns are compatible with both Russian-made chromatographs and instruments from international manufacturers, including Dionex. They perform effectively in both isocratic and gradient modes. Their high resolving power makes them suitable for a wide range of applied and research applications.
Enables the determination of low concentrations of target analytes in the presence of high concentrations of interfering anions.
Specialized applications include the selective determination of oxyhalides, particularly bromates, as well as haloacetates and other carboxylic acid derivatives.
Routine composition analysis of drinking water, groundwater, surface water, and treated wastewater for fluoride, chloride, nitrite, nitrate, sulfate, and other ions using either hydroxide or carbonate eluents.
Supports large injection loops without loss of accuracy when analyzing weakly detectable ions such as fluoride and carboxylates.
Compatible with hydroxide eluents, with detection limits down to hundreds of nanograms per liter.
Compatible with both Russian-made and imported instruments, including Dionex and others.
[4]
[6]
[5]
[3]
[1]
[2]
TECHNICAL SPECIFICATIONS
No. 10 (3/16”) – 32 UNF, tapered
2 µm
PEEK
150 mm × 4.6 mm
Ion-exchange groups:
5 µm
nonporous
styrene-divinylbenzene copolymer
up to 1,000 µL
130 atm
90–100 atm at 1.2 mL/min (15–20 mM NaOH)
compatible
hydroxide, carbonate (pH up to 14)
Eluents:
Aqueous-organic mobile phases:
Operating pressure:
Maximum operating pressure:
Injection loop volume:
Matrix:
Matrix porosity:
Sorbent particle size:
Ion-exchange groups:
Column dimensions:
Column body material:
Inlet/outlet frit:
Inlet/outlet fitting:
TEST DATE:
DATE OF MANUFACTURE:
DIMENSIONS:
13 Feb 2024
15 Jan 2024
150 × 4.6 mm
Hydroxan CE-1
COLUMN:
Certificate No. A101-I06110137
89 atm
50 µL
5 mg/L
5 mg/L
5 mg/L
5 mg/L
5 mg/L
NaF
NaCl
NaNO2
NaNO3
Na2SO4
18 mM NaOH
1.2 mL/min
25 °C
Conductivity detector
Stationary capillary suppressor
Eluent
Flow rate
Temperature
Detector
Suppressor
Injection loop volume
Pressure
SAMPLE COMPOSITION
ANALYSIS CONDITIONS
TEST CHROMATOGRAM
COMPONENTS
640
2440
2930
3600
2,21
1,96
2,03
2,31
6710
-
6,8
11,9
3,2
8,1
1,37
18,55
12,83
5,59
4,43
1,90
fluoride
chloride
nitrite
nitrite
sulfate
1
2
3
4
5
Efficiency, t.plates
Resolution
Asymmetry
Retention time, min
Ion
Peak
STORAGE SOLUTION: 15–25 mM NaOH
Store and transport in the manufacturer’s packaging under dry conditions at +4 to +20 °C, protected from direct sunlight and moisture, and away from heating devices.